Manual RF cryogenic probe station with four RF probe arms and cryogen-free operation
- Pirkėjas
- NPL Management Limited
- Paskelbta
- 2026-10-27
- Terminas
- Nenurodyta
- Šaltiniai
- UK_FTS
- Vertė
- 310000.0 GBP
Aprašymas
Aprašymas pateikiamas šaltinio kalba.
NPL require a RF cryogenic probe station with four installed RF probe arms. The system must support cryogen-free operation using a closed cycle refrigerator. The base temperature must not exceed 5.5 K, and the temperature must be controllable from 6 K to 350 K. The system must support RF measurements from DC to 40 GHz using groundsignal- ground (GSG) probes. The system must accommodate samples with a diameter of at least 2 inches. Vibration at the sample stage must be less than 1 micrometre. The system must include all necessary components, including a turbomolecular pumping system, a microscope with a zoom ratio of at least 12:1, probes, and a calibration substrate. The supplier must install and verify the system and provide user training.
CPV: 33141641, 38000000