Tenderfy

Manual RF cryogenic probe station with four RF probe arms and cryogen-free operation

Acheteur
NPL Management Limited
Publié
2026-10-27
Échéance
Non indiqué
Sources
UK_FTS
Valeur
310000.0 GBP

Description

La description est dans la langue de la source.

NPL require a RF cryogenic probe station with four installed RF probe arms. The system must support cryogen-free operation using a closed cycle refrigerator. The base temperature must not exceed 5.5 K, and the temperature must be controllable from 6 K to 350 K. The system must support RF measurements from DC to 40 GHz using groundsignal- ground (GSG) probes. The system must accommodate samples with a diameter of at least 2 inches. Vibration at the sample stage must be less than 1 micrometre. The system must include all necessary components, including a turbomolecular pumping system, a microscope with a zoom ratio of at least 12:1, probes, and a calibration substrate. The supplier must install and verify the system and provide user training.

CPV: 33141641, 38000000

Abonnement

Consultez gratuitement les aperçus. Les liens vers les avis officiels et leurs documents, les permis de construire et l’accès MCP personnel nécessitent un abonnement payant actif.

Lire